9.
Cleaning and decontamination
9.3.2

UV disinfection using a mobile UV device

To control the disinfection routine of a mobile UV device with the safety
cabinet software, ensure that the connecting plug of the UV device is
compatible with the UV disinfection adapter of the safety cabinet.
To start the UV disinfection routine:
1. Position the mobile UV device centered on the work area and
connect to the UV disinfection adapter.
2. Close the front window completely, the air system operates in the
reduced mode (yellow LED illuminated).
3. Start UV disinfection using the key
Keep the key
UV and the remaining disinfection time.
To interrupt / cancel the UV disinfection routine:
Keep the key
the operating hours) or open the front window.

9.3.3 Changing the UV disinfection time

The UV disinfection time has been preset but can be changed as required:
1. Switch the safety cabinet on and move the front window to the
working position.
2. Keep the key
3. To extend the time: Press the key
extends the time by 30 minutes.
4. To reduce the time: Press the key
). Each pressing of the key reduces the time by 30 minutes.
NATION
5. To save the value: Press the key
operating hours again.
depressed until the display alternatingly shows
depressed until UV is switched off (the display shows
depressed until the preset time is displayed.
. The display indicates the
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
at the operating panel:
. Each pressing of the key
(I
NTERNAL WORK SPACE ILLUMI
55