Installation Test - Thermo Scientific Safe 2020 Operating Instructions Manual

Safety cabinets
Table of Contents

Advertisement

5.
Start-up
5.7

Installation test

Do not operate the device before the installation test has been completed.
The installation inspection of the device must be conducted in accordance
with EN 12469 and additionally according to DIN 12980 for Maxisafe 2020.
The cabinet may be operated as a Class II microbiological safety cabinet,
in accordance with EN 12469 / 2000, if the device functions or function patterns
listed below were checked and if the test results are within the safety value
tolerances specified in Annex F:
Electrical safety test
Inflow velocity test
Downflow velocity test
Leakage test of HEPA filters
Airflow control test
A repeat test must also be performed after repairs to the device or after
major changes (more than 5 cm) to the location of the device.
The operator must prepare a test report or request a written test report from
the authorized test service.
NOTE – Safety warranty
The operational safety of the device,
particularly the personal and material
protection, are guaranteed only if all safety
functions of the device have been tested and
approved.
Thermo Electron LED GmbH will not warrant
the operational safety if the device is operated
without performance of the required
installation test or if the installation test and
repeat test are not performed by adequately
trained and authorized personnel!
NOTE – Device hygiene
The initial start-up with subsequent installation
test does not include any decontamination
measures. For operation in the work process,
the sample chamber of the device and the
accessories required for the work process
must be disinfected and cleaned in accordance
with the hygiene guidelines set forth for the
application.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
31