5.
Start-up
Model Maxisafe 2020:
Fig. 10b: The stand [3] and the prefilter housing form an
assembly. To facilitate the installation of the device frame [1],
the stand is equipped with a lifting mechanism.
The four columns [7] of the stand have threaded rods [6] onto
which the device frame can be placed and lowered.
1. Fig. 10b: Unscrew the threaded rods from their hex disks
[5] by approx. 10 cm and make sure their height is identi-
cal on all four columns.
2. Place the safety cabinet onto the threaded rods. Check to
see whether each guide pin [4] is seated perfectly in its
receptacle at the lower device frame.
3. Screw in each threaded rod at its column in short, even
intervals all the way to the stop of the hex disk.
4. Align the working area of the safety cabinet:
Place a bubble level onto the workplate and rotate the stands
[2] of the stand using a 24-mm wrench until the workplate is
exactly horizontal in all directions. For the vertical alignment
of the device stands, proceed from left to right and from rear
to front.
26
NOTE – Leveling the device
To level the device, do not use the lifting
mechanism but only the adjustable stands of
the device.
Fig. 10b
Stand installation model
Maxisafe 2020