Correct Location; Installation In Line - Thermo Scientific Safe 2020 Operating Instructions Manual

Safety cabinets
Table of Contents

Advertisement

Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
3.
Installation
3.3

Correct location

Choose a draft-free location where the safety cabinet does not
interfere with the room traffic.
Fig. 1: This figure shows preferred locations for safety
cabinets and unsuiable locations, not in accordance with the
safety requirements.
Unsuitable locations: The locations [1], [2], and [3] are not
suitable because they are exposed to drafts from windows and
doors.
Location [5] is undesirable because it is in range of plant traffic
and within the exhaust air range of a ventilation system [4].
Preferred locations [6], [7], and [8] are correct because they
are in a draft-free section of the room and not exposed to plant
traffic.
3.4

Installation in line

When several devices are to be installed in line, please observe
the following:
Make sure that vibrations cannot be transferred between
adjacent units.
Exterior surfaces of the cabinets must always be accessible
for cleaning and disinfection.
12
Fig. 1
Locations in a room

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Maxisafe 2020

Table of Contents