4.
Unit description
Disinfection adapter (optional), Fig. 7a: The disinfection
adapter [9] is used to connect a mobile UV radiation device.
The adapter is connected to the device control, the UV
disinfection routine with a mobile UV device may be controlled
with the remote control.
Media connections (optional): The media supply unit
consists of three pipes that are routed into the sample chamber
through the top of the unit. The inlet connections [1] and
outlet connections [6] with thread (R 3/8"), e.g. for media valves
[7], are preinstalled and equipped with a sealing plug.
The media connections are universal-type connections. Two
equipotential bonding connections are installed at the top of
the unit and at the stand.
Caution – Combustible gas!
If a gas burner is to be operated in the sample
chamber, an appropriate shut-off device for the
gas supply system (shut-off valve, solenoid
valve) must be installed.
Use only laboratory safety burners in the
sample chamber.
External systems: A failure detection systems or gas supply
solenoid valves may be connected to the safety cabinet control.
The unit may also be connected to an external ventilation
system.
4.7

Sample chamber illumination

Fig. 7b: The work space illumination [4] is equipped with two
fluorescent tubes [3] that are installed behind the light dome
[2].
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Fig. 7b
Sample chamber illumination