7.
Operation
7.4

Work rules

The observance of work rules ensures a minimum of operational safety when
handling the safety cabinet.
Before starting an operation:
Take off jewelry.
Put on required personal protective gear, e.g. hand, face, or body protection.
Clean and disinfect sample chamber surfaces at regular intervals.
During operationan:
Place samples only with in the defined work area of the workplate.
Do not place unnecessery items into the sample chamber.
Use only disinfected and cleaned accessories for the work process.
Do not cause air turbulence, by quick hand, arm or body movement in the
sample chamber or in front of the work opening.
Do not place accessories into the sample chamber that cause air turbulence
or emit excessive heat.
Do not block air circulation at the ventilation slots of the workplate.
Sitting posture during work:
Fig. 16: To prevent risks to health, a height-adjustable working
chair with an adjustable seat back should be used during
extended work periods at the safety cabinet.
A When the forearm rests on the armrest, is should be in an
almost horizontal position.
B When the thigh is in a horizontal position, the angle between
thigh and lower leg should exceed 90°.
To ensure a compensation between floor and sitting height, a
Footrest (optional accessory) should be used.
After finishing an operation:
Remove samples from the sample chamber and store them
properly.
Clean and disinfect the sample chamber surfaces,
including the workplate and the floorpan. Clean and
disinfect all accessories.
Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
Fig. 16
Sitting posture
51