Extending The Output Power; Series Connections; Figure 2-13 U2722A/U2723A Outputs Connected In Series - Keysight U2722A User Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

Extending the Output Power

The U2722A/U2723A is equipped with three dedicated channels of up to 20 V and
120 mA per channel.

Series connections

To increase the output voltage, tie two or more of the outputs in series as shown
in
output. Series connected outputs can be operated with one load across the
connected outputs or with a separate load for each output.
When a series connection is used, the output voltage is the sum of the voltages of
the individual output. The current is the current of any one output. Each of the
individual outputs must be adjusted to obtain the total output voltage.
Positive (+) terminal of the device-under-test
Figure 2-13
– Ensure that the voltage (when source current) or the current (when source
NOTE
– Preferably, both are at the same current range.
Keysight U2722A/U2723A User's Guide
Figure
2-13. This is to obtain a higher voltage than that available from a single
Output 1
SENSE OUTPUT GUARD OUTPUT SENSE SENSE OUTPUT GUARD OUTPUT SENSE SENSE OUTPUT GUARD OUTPUT SENSE
+
+
-
U2722A/U2723A outputs connected in series
voltage) is not clamped.
Output 2
-
+
+
Negative (–) terminal of the device-under-test
Operation and Features
Output 3
-
-
+
+
2
-
-
53

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2723a

Table of Contents