Error Conditions; Scpi Commands For System-Related Tasks - Keysight U2722A User Manual

Usb modular source measure units
Hide thumbs Also See for U2722A:
Table of Contents

Advertisement

2
Operation and Features

Error conditions

A record of up to 20 errors can be stored in the U2722A/U2723A's error queue.
Refer to the programming guide for more information on the error messages.
Keysight Measurement Manager operation
A message box will appear once an error occurs while operating the U2722A/
U2723A using the Keysight Measurement Manager.

SCPI commands for system-related tasks

The following examples show the SCPI commands for performing certain
system-related tasks.
Example 3, Performing system-related tasks
→ *CLS; *RST
→ *TST?
← +0
→ *CAL?
← +0
→ SYST:ERR?
← +0, "No Error"
52
Resets the source to the default
power-on state. This command can be
ignored if this operation is not required.
Executes the self-test.
Returns a +0 if the test pass, else it
returns other numbers in correspond
with the failure (refer to
List of Self-Test Return Codes"
page 66).
Executes the self-calibration.
Returns a +0 if the test pass, else it
returns a +1 if it fails.
Returns the error number and its
corresponding message string from the
error queue.
Keysight U2722A/U2723A User's Guide
"Appendix A:
on

Advertisement

Table of Contents
loading

This manual is also suitable for:

U2723a

Table of Contents