2
Operation and Features

Parallel connections

A parallel connection is used to increase the output current. Connect two or more
of the outputs in parallel as in
current is the sum of the currents of the individual output. The voltage is from any
of the outputs. Each of the individual outputs must be adjusted to obtain the total
output current.
Positive (+) terminal of the device-under-test
Figure 2-14
– Ensure that the voltage (when source current) or the current (when source
NOTE
– Preferably, both are at the same voltage range.
54
Output 1
SENSE OUTPUT GUARD OUTPUT SENSE SENSE OUTPUT GUARD OUTPUT SENSE SENSE OUTPUT GUARD OUTPUT SENSE
+
+
-
U2722A/U2723A outputs connected in parallel
voltage) is not clamped.
Figure
2-14. For a parallel connection, the output
Output 2
-
+
+
Negative (–) terminal of the device-under-test