Command Syntax
Parameters
Examples
Related Commands
INITiate:CONTinuous:SEQuence
INITiate:CONTinuous:NAME
These equivalent commands prepare the list to respond to trigger commands. ON or 1 continuously
initiates the list. OFF or 0 turns off continuous initiation. Upon the receipt of a trigger with continuous
initiation on, one of the following actions occur:
If LIST:STEP is set to ONCe, the list will progress to the next step in the sequence.
If LIST:STEP is AUTO, each trigger will start the list again.
These commands are not channel specific, they apply to the entire mainframe. If the trigger system is not
in the Idle state and therefore already initiated, the initiate commands are ignored.
Command Syntax
Parameters
Examples
Related Commands
TRIGger
When the trigger system has been initiated, this command generates a trigger signal regardless of the
selected trigger source. This command is not channel specific, it applies to the entire mainframe.
Command Syntax
Parameters
Examples
Related Commands
TRIGger:DELay
Channel Specific
This command sets the time delay between the detection of a trigger signal and the start of any
corresponding trigger action. After the time delay has elapsed, the trigger is implemented. This command
only applies to the selected channel.
Command Syntax
Parameters
*RST Value
Examples
Query Syntax
Returned Parameters
Related Commands
INITiate[:IMMediate]:SEQuence2
INITiate[:IMMediate]:NAME ACQuire
For INIT:NAME: ACQuire
INIT:SEQ2
ABOR
INIT:CONT
INITiate:CONTinuous:SEQuence[1] <bool>
INITiate:CONTinuous:NAME LIST, <bool>
0 | 1 | OFF | ON
INIT:CONT:SEQ1 ON
ABOR
INIT:CONT
TRIGger[:IMMediate]
None
TRIG
ABOR
TRIG:SOUR
TRIGger:DELay <NRf+>
0 - 0.032s | MINimum | MAXimum
seconds
Unit
0
TRIG:DEL .025
TRIGger:DELay?
<NR3>
ABOR
TRIG
INIT:NAME ACQ
TRIG
*TRG
INIT:CONT:NAME LIST, 1
TRIG
*TRG
TRIG:DEL
TRIG:TIM
TRIG:DEL MAX
TRIG:SOUR
TRIG:TIM