Methods 1–150 are predefined methods. The analytical conditions are predefined. Some
of the fields of the analytical conditions are grayed out and cannot be edited.
These are Siemens defined methods, only edit user-defined parameters.
NOTE:
Methods 151–200 are reserved for user-defined methods.
When you change and save your settings in the analytical parameter fields the
NOTE:
version number in the upper-right corner of the window moves up one numerical
increment and the date is updated to the current date, for example, from Ver.
2.0.0.yyyymmdd to Ver. 2.0.1.yyyymmdd (in this example the 2 is the software version
and the first 0 is the method parameter revision minor number).

System Test List window

Perform any of the following tasks at this window:
Add tests that are used regularly by your system. The tests on the list appear on other
windows
Assign a reagent tray position number for non-barcoded reagents or a reagent code
(R-Code) for barcoded reagents
CAUTION
To avoid incorrect reporting of results at the Review and Edit, the Daily QC
windows, and the Print Report form, you
items on the System Test List window after initial setup. If you reposition any test
items on the System Test List window, results for samples already run could be
associated with the wrong test name.
Perform any of the following tasks at this window:
add tests to list
remove tests from list
set the test processing order
Analysis order oet window
Use this window to set the order in which tests on the list run. To run tests in the order
they displays at the System Test List window, you do not need to use this window.
print test list items
assign reagent codes or position numbers

Analytical Parameters (Serum) window*

Use this window to set the parameters for evaluating the serum indices (lipemia,
hemolysis, and icterus).
Perform any of the following tasks at this window:
define up to five different flagging levels of concern for each serum indices
select a test item to perform the serum indices analyses
enter the factor values required to calculate the serum indices result values
Operation of the serum indices feature is controlled in the Serum set area of the
NOTE:
System Specifications Set window.