Every 2 Months Maintenance; Cleaning The Dilution Tray Cuvettes - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

3. Locate the filter and slide it out of the analyzer.
4. Using a vacuum cleaner, remove the dust from the filter.
5. If the filter requires further cleaning:
a. Wash the filter under running water.
b. Dry the filter before replacing it.
6. Slide the filter back in place and close the panel on the right side of the cabinet.
7. Return the analyzer to the READY mode.

Every 2 months maintenance

Cleaning the dilution tray cuvettes

Materials required:
• 2-liter beaker
• Probe Wash 3 solution (REF 03164495,
PN B01-4183-01)
• Deionized water
Time: 15 minutes (replacement)
10 hours (immersion)
Analyzer mode: Off
Figure 5-19. Components of dilution tray cuvette
1. Prepare 1.5 liters of 5% Probe Wash 3 solution diluted with deionized water.
2. Remove the 6 cuvette segments on the dilution tray (DTT).
a. Unfasten the 2 thumbscrews (1) on each section.
b. Grasp the cuvette section by the tab (2) and lift it from the tray.
c. To remove the cuvettes under the dilution washer (DWUD) and splash cover,
Maintenance
turn the tray by hand until the cuvettes are clear.
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
131