Cleaning And Replenishing The Cuvette Conditioner Bottle - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

3. Immerse the cuvette segments in 5% Probe Wash 3 solution.
a. Ensure no air bubbles are in the cuvettes.
b. Allow the cuvettes to soak for at least 10 hours.
4. Wash the cuvettes under running water, then rinse them in deionized water.
5. Drain the water from the cuvettes.
6. Install the cuvette segments on the dilution tray (DTT) and fasten the thumbscrews
by hand.
7. At the Operation Panel, select Initialize.
NOTE:
further actions.
8. At the Operation Panel, select WASH.
9. Perform the daily shutdown wash routine, then verify the operation.

Cleaning and replenishing the cuvette conditioner bottle

Materials required:
• Deionized water
• Cuvette conditioner
Time: 10 minutes
Analyzer mode: READY
NOTE:
cleaned at least once every 2 months.
Figure 5-20. Cuvette conditioner bottle
1. Open the filter cap at the front of the cuvette conditioner bottle (
tube with the filter.
2. Disconnect the cuvette conditioner bottle level sensor connector.
3. Turn the connector counter-clockwise and pull it out.
132
Verify that the Operating mode field displays READY before performing any
The cuvette conditioner bottle may be cleaned when it is refilled, but must be
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
1)
ISE buffer bottle
2)
RRV (Reaction) bath oil
3) Isotonic saline diluent bottle
4) Cuvette detergent bottle
5) Cell conditioner bottle
), and pull up the
5
ADVIA 2400 Operator's Guide