Performing Additional Ise Electrode Washes - Siemens ADVIA 2400 Operator's Manual

Chemistry system
Hide thumbs Also See for ADVIA 2400:
Table of Contents

Advertisement

6. At the WASH Set window, define the WASH2 container positions as follows:
a. Select
b. Select
c. Enter
d. Enter
e. Select

Performing additional ISE electrode washes

Materials required:
• ISE Detergent Solution
(REF 01307361, PN B01-4174-01)
Time: 5 minutes
Analyzer mode: Manual Operation
118
.
WASH2
for Cycles.
2
in the CTT cup position 1st time field and
49
2nd time field.
in the RTT1 and RTT2 cup positions 1st time fields and
49
and RTT2 cup positions 2nd time fields.
.
Execute
CAUTION
Use only Siemens-recommended
ancillary reagents to avoid
erroneous test results.
in the CTT cup position
50
BIOHAZARD
Wear personal protective
equipment.
Use universal precautions.
WARNINGS
• Read and follow the cautions on
the box and label before
handling any reagents. Some
method and system reagents are
classified as hazardous material.
• Avoid contact of method or
system reagents with skin and
eyes. These materials can cause
infection and burns. Wear
suitable protective clothing,
gloves, and eye/face protection.
In case of contact with eyes,
rinse immediately with plenty
of water and seek medical
advice. In case of contact with
skin, immediately wash with
soap and water.
• Avoid inhaling chemical
vapors. If you inhale chemical
vapors, promptly leave the area
and seek fresh air.
ADVIA 2400 Operator's Guide