Emerson Fisher 4320 Instruction Manual page 4

Wireless position monitor
Hide thumbs Also See for Fisher 4320:
Table of Contents

Advertisement

4320 Position Monitor
February 2013
Table 1. Specifications (continued)
Electrical Classification
CSA (C/US)— Intrinsically Safe
ATEX— Intrinsically Safe
IECEx— Intrinsically Safe
Refer to Hazardous Area Classifications and Special
Instruction for "Safe Use" and Installations in
Hazardous locations, starting on page 11, for
additional information
Electrical Housing
4X & IP66
Other Classifications/Certifications
INMETRO— Brazil
NEPSI— China
Contact your Emerson Process Management sales
office for classification/certification specific
information
IEC 61010 Compliance
Meets Pollution Degree 2
Adjustments
Zero and Maximum Span through local interface
Weight
0.5 Kg (1.1 lbs)
(2)
Power Module
Standard —Lithium, non‐rechargeable
1. The pressure/temperature limits in this manual and any applicable standard or code limitation for valve should not be exceeded.
2. Due to the combustible nature of the lithium content, the power module has special installation, operation, storage, and/or shipping requirements. Observe all warnings included with the power
module before installing, operating, storing, or shipping. Contact your Emerson Process Management sales office if additional information is needed.
Table 2. EMC Summary Results—Immunity
Port
Phenomenon
Electrostatic discharge (ESD)
Radiated EM field
Enclosure
Rated power frequency
magnetic field
Burst
(3)
Surge
DC Power
Conducted RF
Performance criteria: +/- 2% of effect
1. A = No degradation during testing. B = Temporary degradation during testing, but is self‐recovering.
2. Battery Sourced Power Module = B ; External Power Option = A
3. External Power Option only.
4
Power Module Limits
Standard— 10 years at update rate of 16 seconds or
longer with 3 additional devices communicating
through it
Extended‐Life— 10 years at update rate of 4 seconds
or longer with 3 additional devices communicating
through it
Shelf life— 10 years (radio off)
External Power (Optional)
12 - 28 volts DC
15 mA maximum operating current
Polarity Insensitive
Wire Size— 14-20 gauge
Update Rates
For position feedback, temperature, and power
module voltage—selectable from 1 second to 1 hour;
1s, 2s, 4s, 8s, 16s, 32s, 1 to 60 minutes
Factory default is 1 minute
Note: Sample rate for position feedback is normally
the same as the update rate. When using Report by
Exception reporting refer to Advanced Wireless
Reporting, on page 27, for additional information.
Construction Material
Housing: Engineered resin
Basic Standard
6 kV contact
IEC 61000‐4‐2
8 kV air
80 to 1000 MHz @ 10 V/m with 1 kHz AM at 80%
1400 to 2000 MHz @ 10 V/m with 1 kHz AM at 80%
Battery Sourced Power Module
IEC 61000‐4‐3
2000 to 2700 MHz @ 3 V/m with 1 kHz AM at 80%
External Power Option
2000 to 2700 MHz @ 1 V/m with 1 kHz AM at 80%
IEC 61000‐4‐8
100 A/m @ 50 Hz & 60 Hz
IEC 61000‐4‐4
2 kV (5/50 ns, 5 kHz)
IEC 61000‐4‐5
1 kV line to line / 2 kV line to ground
IEC 61000‐4‐6
10 Vrms (150 kHz to 80 MHz)
Instruction Manual
Test Level
D103392X012
(1)
Performance Criteria
(2)
A
A
A
A
A

Advertisement

Table of Contents
loading

Table of Contents