OXIMETER
Oxygen Saturation Range
(%SpO
)
2
Pulse Rate Range
Displays
Numeric Displays
Pulse Strength
Bargraph LED
Saturation Accuracy
(%SpO
) (±1 S.D.)*
2
No Motion
Adults, Pediatrics
Neonates
Motion
Adults, Pediatrics
Neonates
Low Perfusion
Adults, Pediatrics
Neonates
Pulse Rate Accuracy
No Motion
Adults, Pediatrics,
and Neonates
Motion
Adults, Pediatrics,
and Neonates
Low Perfusion
Adults, Pediatrics,
and Neonates
Measurement Wavelengths and Output Power
Red
Infrared
Altitude
Operating Altitude
* S.D. (standard deviation) is a statistical measure; up to 32% of readings may fall outside these
limits.
31

Specifications

0% to 100%
18 to 300 pulses per minute
3-digit LEDs, Tricolor (red, green, amber)
Tricolor LED segments
70-100%
8000AA
Finger Clip
Sensor
±2 digits
±3 digits
±2 digits
±3 digits
±2 digits
±3 digits
18 to 300 beats/min. (no motion)
60 to 240 beats/min. (motion)
8000AA
Finger Clip
Sensor
±3 digits
±5 digits
±3 digits
660 nanometers @ 3 mw nominal
910 nanometers @ 3 mw nominal
Up to 40,000 feet
8000J
Flex Sensor
±2 digits
±3 digits
±2 digits
±3 digits
±2 digits
±3 digits
8000J
Flex Sensor
±3 digits
±5 digits
±3 digits