Specifications - Nonin 8600V Operator's Manual

Veterinary pulse oximeter
Table of Contents

Advertisement

Specifications

1. Oxygen Saturation Range
(SpO
)
2
2. Pulse Rate Range
3. Displays
Animal Indicators
Digital Displays
Equipment Indicators
4. Measurement Wavelengths
Red
Infrared
5. Accuracy ∗
SpO
2
(± 1 Standard Deviation) ♦
Pulse Rate
6. Alarm Ranges
High SpO
limit
2
Low SpO
limit
2
High pulse rate limit
Low pulse rate limit
7. Alarm Volume
8. Temperature
Operating
Non-operating
9. Humidity
Operating
Non-operating
10. Battery Charger
Model 8604A
Model 8000A
11. Batteries
13. Dimensions
14. Weight
♦ Standard Deviation is a statistical measure; up to 32% of the readings may fall outside these limits.
∗ Accuracy is specified for adult human hemoglobin measured at the fingertip. Other types of hemoglobin may affect accuracy.
0 to 100%
18 to 450 pulses per minute
Perfusion, SpO2 Low, SpO2 High
Pulse Fast, Pulse Slow
3-digit 7-segment LEDs
Sensor, Self Test, Battery Low, Battery Charging, and
Audible Alarm Disabled indicators
660 nanometers
910 nanometers
70 - 100% ± 2 digits
Below 70% is not specified for all sensors
± 3% ± 1 digit
85%, 90%, 92%, 94%, 95%, 96%, 97%, 98%, 99%, OFF
55%, 60%, 65%, 70%, 75%, 80%, 85%, 86%, 87%, 88%, 89%, 90%, 91%, 92%, 95%, OFF
60BPM, 75BPM, 100BPM, 125BPM, 150BPM, 175BPM, 200BPM, 225BPM, 250BPM, 275BPM,
300BPM, 325BPM, 350BPM, 375BPM, 425BPM, OFF
20BPM, 25BPM, 30BPM, 40BPM, 50BPM, 60BPM, 70BPM, 80BPM, 100BPM, 120BPM, 140BPM,
160BPM, 180BPM, 200BPM,250BPM, OFF
Adjustable to 85 dbA at 2 ft. (60 cm)
0°C to +45 °C (+32°F to +113°F)
-20°C to +45 °C (-4°F to +113°F)
10 to 90% non-condensing
10 to 95% non-condensing
120 VAC / 60 Hz
230 VAC / 50 Hz
5-cell rechargeable Ni-Cad battery pack
30 hours minimum operation
15 hours to fully recharge
5.8" wide x 2.7" high x 7.5" deep
14.7 cm x 6.9 cm x 19 cm
2 lbs (0.75 kg)
24

Advertisement

Table of Contents
loading

This manual is also suitable for:

8600mv

Table of Contents