8.2 Error – Cause – Remedy
8.2.1 Pop‑up message
Error
` Possible cause
Communica‑
One or more
`
`
VACUU·BUS compo‑
tion error
nents were removed.
Error at
Address incorrectly
`
`
configured.
frequency
converter (FC)
Temperature too
`
`
high.
FC defective.
`
`
Error at control
Valve defective.
`
`
system
Error at pump
Check VMS‑B
`
`
(switching device).
Error at digital
No power supply at
`
`
I/O module
IN of I/O module.
Plug pulled out.
`
`
An error occurred in
`
`
the system and the
I/O module relayed it
to the controller.
Error at analog
No power supply.
`
`
I/O module
Error at
Ambient temperature
`
`
too high, Peltronic
Peltronic
overheated.
Performance require‑
`
`
ments too high.
Peltronic defective.
`
`
Error at vacu‑
Vacuum sensor
`
`
defective.
um sensor
Overpressure
Pressure too high.
`
`
Measuring range
`
`
exceeded.
Underrange
Pressure below mea‑
`
`
suring range.
Vacuum sensor
`
`
adjustment incorrect.
Troubleshooting
Remedy
Deactivate relevant
`
VACUU·BUS com‑
ponents.
Perform component
`
detection.
Configure correct
`
address.
Replace defective
`
component.
Check address.
`
Replace defective
`
component.
Send in defective
`
device.
Connect power sup‑
`
ply.
Check plug‑in con‑
`
nection.
Remedy cause of
`
external error.
Connect power sup‑
`
ply.
Eliminate cause of
`
overheating of the
Peltronic.
Send in defective
`
component.
Replace defective
`
component.
Send in defective
`
component.
Acknowledge warn‑
`
ing indication.
Eliminate cause of
`
overpressure.
Calibrate vacuum
`
sensor correctly.
Personnel
Specialist
Resp. spe-
cialist
Specialist
Resp.
specialist
Specialist,
resp.
specialist
Specialist
Specialist
Resp.
specialist
Operator,
specialist
Specialist
75