vacuubrand VACUU-SELECT Instructions For Use Manual page 65

Vacuum controller
Hide thumbs Also See for VACUU-SELECT:
Table of Contents

Advertisement

 Example
Overview
Context menu
settings
Overview of
possible basic
settings
20901059_EN_VACUU·SELECT_V1.3_280119
Meaning of basic settings
Function
Autostart
Vacuum sensor
Run‑on time for cool‑
ant valve(s)
Delay time for level
sensor(s)
The available basic settings adapt to the connected VACUU·BUS
components.
You can specify presettings for your
process in Basic Settings. The
vacuum sensor, which measures
actual pressure, is shown here.
Setting
Meaning
Off / On
Off: The controller is set to
"stop" after power is restored.
On: An application that was
started is continued following a
power failure and subsequent
restoration of power.
VS‑C _ / VS‑P _
Select the vacuum sensor for
control. VSK: rough vacuum,
VSP: fine vacuum.
Off / hh:mm:ss
Coolant run‑on time.
Off / hh:mm:ss
Delay time for full status indi‑
cator of level sensor.
Main menu
65