20901059_EN_VACUU·SELECT_V1.3_280119
Error
` Possible cause
Transfer failed
No USB flash drive
`
`
connected.
Not enough storage
`
`
space on the USB
flash drive.
Venting valve
No voltage applied.
`
`
does not oper‑
VACUU·BUS plug‑in
`
`
ate
connection or cables
defective or not con‑
nected.
Venting valve dirty
`
`
(polluted).
Venting valve in sen‑
`
`
sor defective.
Venting valve deacti‑
`
`
vated.
Troubleshooting
Remedy
Connect a USB
`
flash drive with
sufficient storage
space.
Check VACUU·BUS
`
plug‑in connection
and cables to the
controller.
Clean venting valve.
`
If necessary, use
`
another external
venting valve.
Activate venting
`
valve in the control‑
ler.
Personnel