Installation and connection
(b)
(a)
(c)
(d)
(b)
(a)
36
Sensor connection via small flange
Required connection material: VACUU·BUS extension cable
for connection to the controller (option), clamping ring with uni‑
versal centering ring or inner centering ring for KF DN 16 (tool:
open‑end wrench SW17).
1. Remove the blind plug and place the small flange KF DN 16
(b) on the vacuum connection of the sensor (a).
2. Tighten the small flange KF DN 16 until hand‑tight.
3. Remove the protective dust cap (c).
4. Place the sensor with the centering ring on the connection of
the equipment small flange KF DN 16 (b).
5. Secure the sensor (a) with the clamping ring (d) to the vacu‑
um line, as shown in the illustration.