MAINTENANCE AND TROUBLESHOOTING

Overview

AVTM246004 BITE 2/2P Rev 4 July 2011
Find Quality Products Online at:
This chapter explains:
How to maintain the BITE 2/2P so that it remains in
good operating condition
How to interpret error messages
What to do if repairs are needed
NOTE: Before performing maintenance on the BITE 2 or BITE
2P, please read, understand, and observe all safety precautions
as indicated in Chapter 2, "Safety". Maintenance should be
performed only by qualified personnel familiar with the hazards
involved with line-operated test equipment.
The BITE 2/2P transmitters contain large high-voltage
capacitors. During operation these capacitors could be
charged up to about 300 V dc. Normally these capacitors
F
are automatically discharged when the instrument is
switched off. However, under certain fault conditions,
these capacitors may be left charged. Always use a
voltmeter to check the state of the charge and wear
rubber gloves as necessary when touching the
capacitors and the circuits connected to them.