Table 14 - Instrument Configuration Parameters
Parameter
Description
Multi Meas
Determines what is viewed on the screen while taking
View
multiple measurement location.
Trigger
Sets the trigger source.
Trig Level
Sets the trigger level. Applies only to triggered measurements (rpm points, phase points, time-synchronous averaged points).
Value
Automatic
Fixed: Default The Dynamix 2500 data collector uses the signal level entered in the point setup information (either
Hist. Data
Determines how data is stored in the Dynamix 2500 data
Collect
collector when collecting multiple measurements for a signal
point:
• If you want to conserve space, you should select Always
Overwrite.
• If you want multiple data sets to be stored and unloaded,
select Always Append.
Rockwell Automation Publication 1441-UM001B-EN-P - September 2012
Description
Samples the trigger signal before collecting the triggered measurement and sets an appropriate trigger
level automatically. When collecting data on a triggered point, a message displays on the screen while
the trigger level is being sampled.
The length of time for which the trigger is sampled during autotriggering is dependent on two factors:
• The high pass filter selected for the point.
• The lower the high pass filter cutoff, the longer the trigger will be sampled.
Mach. Speed
If the Mach. Speed is set to `Normal,' the maximum trigger sampling time is about one second. If the
Mach. Speed is set to `Slow,' the maximum trigger sampling time is about 10 seconds. The Normal
setting supports autotriggering on machines running at speeds down to ~120 rpm. The Slow setting
will support autotriggering on machines running at speeds down to ~6 rpm, depending on the width of
the trigger pulse.
Route or Offroute) as the trigger point.
Configuring the Dynamic 2500 Data Collector
Value
First (default)
Last
All
None
External (default)
Uses the trigger input from the POWER/USB/TRIGGER
connector as the trigger source.
Laser Tach
Enables the internal Laser Tachometer when a Phase or Order
measurement is collected.
Always Append: adds the new data for a point to existing
data.
Always Overwrite: replaces existing data with new data.
Always Ask (default): displays a dialog box before attempting
to store the data.