Configuring Automatic Adjustment Of The Fid - ABB ACF5000 Manual

Multi-component ftir analyzer system
Table of Contents

Advertisement

86

Configuring automatic adjustment of the FID

Function
Menu path
Display
Procedure
Parameter
Enable
Cycle time
Date/time of next
adjustment
ACF5000 FTIR ANALYZER SYSTEM
For the FID, an automatic adjustment with test gas at the zero and end point
can be configured.
Menu → Configure → Adjustment Data →
Automatic Adj.
The recommended settings are stated in the following table.
Explanation
Automatic adjustment is completed only when it is activated.
The cycle time shows the time intervals over which automatic adjustment is
completed. Recommended setting: 21 days
The analyzer system completes the next automatic adjustment at the time
specified here. From this moment in time, the cycle period starts to run.
|
OI/ACF5000-EN REV. A