No.
Text
116
The Profibus module is in-
stalled on a wrong slot! The
Profibus interface is therefore
not functional. Please install
the Profibus module on the slot
X20/X21.
117
The configuration backup was
saved.
118
The configuration backup was
loaded and the system re-
started.
119
The system configuration could
not be loaded. For that reason,
the system does not currently
contain a configuration. Please
load in the menu: Config-
ure/System/
Save configuration the backup
configuration. Or load a con-
figuration with the use of SMT.
120
All passwords are reset to the
system defaults. Please save the
configuration, power off the
system and remove the jumper.
Then repower the system.
121
The maintenance switch is ON.
122
The system disk has been rei-
nitialized! This system contains
no configuration now! Please
use SMT to reinstall your con-
figuration.
QAL3
126
The QAL3 historien buffer is
full. Please read-out QAL3 data.
127
The calibration drift exceed the
QAL3 limits.
System bus
201
The system bus module se-
lected could not be found.
203
The system bus module does
not exist.
208
The system bus could not
transfer any data to the data-
base.
209
The system bus connection
to this module is interrupted.
210
The configuration of the sys-
tem bus module has changed.
211
The system bus module no
longer has an internal memory.
ACF5000 FTIR ANALYZER SYSTEM
S A W F Reaction/Comment
x x
see message text
x x
see message text
see message text
x for information
x
see message text
x
see message text
x
x x
Check plug connections and terminal resistances on the
system bus. Check whether the serial number of the
system bus was entered correctly.
MENU → Diagnosis/Information → System overview
x x
Check plug connections and terminal resistances on the
system bus.
x x
The software version of the system bus module is not
compatible with that of the system controller; carry out
system update of the system controller.
x x
Check system bus connection to the system bus module
displayed. Check power supply of the displayed system
bus module.
x x
for information; the configuration data are updated
automatically
x x
Check configuration of the system bus module:
MENU → Diagnosis/Information → System overview
|