184
No.
Text
399
The measured value exceeds
the value range of the ana-
log/digital converter.
400 A calculation error occurred
while calculating the measured
value.
401
The flow exceeds or falls below
the top or bottom limit value 1.
402
The flow exceeds or falls below
the top or bottom limit value 2.
403
The position value of the flow
controller is outside the valid
area.
FID
411
The analyzer is in standby. Re-
activation in the menu: Service/
Test..Standby/Restart FID
412
Ignition failed. The analyzer
needs to be reactivated manu-
ally. Reactivation in the menu:
Service/Test..Standby/Restart
FID
413
Failure of auxiliary current in
the analyzer hardware.
414
The position value of this con-
troller is below the permissible
range. (< 20%)
415
The position value of this con-
troller is above the permissible
range. (> 90%)
Adjustment
500 System bus communication
disrupted.
501
Required functionality is not
available in the system module.
502
A system error occurred in the
system module addressed.
503
Amplification error during ad-
justment. Adjustment impossi-
ble.
507
A combination of the following
errors has occurred: drift half,
drift, amplification of the delta
drift.
508 Unknown error number. Check
software versions.
509 Automatic adjustment started.
510
Automatic adjustment ended.
511
Automatic adjustment canceled
externally.
ACF5000 FTIR ANALYZER SYSTEM
S A W F Reaction/Comment
x x
Check measuring gas concentration. Check push con-
nections in the analyzer system Notify service.
x x
Notify service.
x
Check measuring gas path. Notify service.
x x
Check measuring gas path. Notify service.
x x
Notify service.
x
x Restart FID.
x x
Check operating gases.
Restart FID.
x x
Notify service.
x
x Check operating gases and connection leads. Restart
FID, where applicable. Notify service.
x
x
Check software version of the analyzer module and carry
out update, if applicable.
Adjustment is canceled.
Notify service.
x
Adjustment is canceled.
End point gas concentration too low – check.
Adjust detector displayed manually at the zero point and
end point.
Message during automatic adjustment. Check software
version of analyzer module and system controller.
for information
for information
for information
|
OI/ACF5000-EN REV. A