28
Fuses
Function
–F80
Supply ground-fault circuit interrupter
–F81
Fan or
air conditioner (option)
–F82
Lighting, service socket
–F83
Type 42 probe tube heater, PFE2 filter unit heater,
back-purging valves
–F84
Heated sample gas line
–F85
ACF5000 electronics box AU1, ASP block heater, gas cell
heater
–F86
Air purifier, FTIR spectrometer, flow monitor, system
controller, power supply unit 24 V/5 A
–F87
Option "2nd measuring location": Type 42 probe tube
heater, PFE2 filter unit heater, back-purging valves
–F88
Option "2nd measuring location": heated sample gas line
–F89
Option "2nd measuring location": heated selector valve
and heated sample gas line to analyzer cabinet
–F90
UPS supply ground-fault circuit interrupter
–F91
Relay coils, contactor coils, solid-state relays, selector
to
solenoid valve (ceramic fuses)
–F99
ACF5000 FTIR ANALYZER SYSTEM
High leakage current: 9 mA!
|
OI/ACF5000-EN REV. A
Rated value
30 A/30 mA
6 A or
16 A at 230 V, 20 A at 120 V
6 A
6 A (only PFE2) or
10 A (PFE2 + probe 42, 230 V) or
16 A (PFE2 + probe 42, 120 V)
16 A
6 A
6 A
6 A (only PFE2) or
10 A (PFE2 + tube 42, 230 V) or
16 A (PFE2 + tube 42, 120 V)
16 A
6 A
25 A/30 mA
T 2 A