No.
Text
328
A calculation error occurred
while calculating the measured
value.
329
The pressure exceeds or falls
below the top or bottom limit
value 1.
330
The pressure exceeds or falls
below the top or bottom limit
value 2.
331
The position value of the pres-
sure is outside the valid area.
I/O devices
332
Loss of auxiliary current in the
I/O card.
333
An I/IO type that is not yet
available is configured.
334
No new measured values from
the analog/digital converter.
335
The measured value exceeds
the value range of the ana-
log/digital converter.
336
A calculation error occurred
while calculating the measured
value.
337
Line break and analog output.
338
Line break in digital input (hu-
midity sensor).
339
Line break or short-circuit in
the analog input.
340
The value of the analog input
exceeds or falls below the top
or bottom limit value 1.
341
The value of the analog input
exceeds or falls below the top
or bottom limit value 2.
Flow monitor
342
The flow rate undershoots limit
value 1
343
Flow rate undershoots limit
value 2.
Measured value
344
The measured value exceeds
the value range of the measur-
ing range.
345
The measured value falls below
the value range of the measur-
ing range.
Flow controller
398
No new measured values from
the analog/digital converter.
ACF5000 FTIR ANALYZER SYSTEM
S A W F Reaction/Comment
x x
Notify service.
x
FID: Check supply gas pressures:
Output = instrument air,
Air = combustion air,
H2 = combustion gas.
x
x x
FID: Check supply gas pressures.
x x
The I/O card is faulty. Replace card.
x x
Correct configuration with test and adjustment soft-
ware.
x x
The I/O card is faulty. Replace card.
x x
Check signals on the analog inputs.
If OK, check configuration and adjustment of the analog
inputs.
x x
Check configuration and adjustment of the analogies
and outputs.
x x
Check lines in analog output.
x x
Check humidity sensor in the system cooler.
x x
Check temperature of the system cooler.
x
Check temperature of the system cooler.
x
Check temperature of the system cooler.
x
Check sample preparation.
Alarm value 1 = 25 % of MRS.
x x
Check sample preparation.
Alarm value 2 = 10 % of MRS.
Automatic calibration is interrupted and disabled
Measured value > +130% MRS; not in the logbook
Measured value < −100% MRS; not in the logbook
x x
Notify service.
|
OI/ACF5000-EN REV. A