ABB ACF5000 Manual page 82

Multi-component ftir analyzer system
Table of Contents

Advertisement

82
Settings
Parameter
Current number of data
entries
Maximum number of data
entries
Display warning when
percentage reached is
ACF5000 FTIR ANALYZER SYSTEM
Whether the analyzer system transmits a status signal according to the limit
values criteria on the CUSUM control chart or the Shewhart control chart, or
according to both control charts, can be selected in the Settings →
Available options for QAL3 control chart submenu.
The number of lines shown in the results table in the Measurements sub-
menu can be set in the Settings → Display submenu. The number of lines
for the first page and the following pages can be set separately.
The storage properties for the QAL3 report can be configured in the Set-
tings → Data storage submenu. The size of the QAL3 report can be
defined (number of lines). If the maximum number of lines is reached, the
oldest entries are overwritten.
Explanation
Displays the current number of data entries.
The maximum number of lines for the QAL3 report must be set.
A storage level must be set (a % of the maximum number) that will trigger a
warning when it is reached.
All previous recorded QAL3 results tables are deleted with the Delete Data
key. As this process cannot be undone, it is recommended that all results
tables are exported before proceeding.
|
OI/ACF5000-EN REV. A