No.
Text
214
The system is currently being
maintained with Optima SMT.
215
The analyzer module has an
internal communication error!
216
The analyzer module has an
internal program error!
250
The analyzer module could not
be found.
251
The connection to the analyzer
module was lost.
252
The EEPROM data of the ana-
lyzer are faulty.
253
The communication with the
analyzer is disrupted,
254
The boot program of the ana-
lyzer module is faulty. Notify
Service!
255
The program of the analyzer
module is faulty. Notify Service!
Analyzer modules
300 No new measured values from
the analog/digital converter.
301
The measured value exceeds
the value range of the ana-
log/digital converter.
302
The offset drift exceeds the
half of the range permissible.
303
The offset drift exceeds the
permissible range.
304
The amplification drift exceeds
the half of the range permissi-
ble.
305
The amplification drift exceeds
the permissible range.
306
The offset drift between two
adjustments exceeds the per-
missible range.
307
The amplification drift between
two adjustments exceeds the
permissible range.
308
A calculation error occurred
while calculating the measured
value.
309
The thermostat works errone-
ously.
310
The temperature correction for
this component was deac-
tivated because the tempera-
ture measured value is invalid.
ACF5000 FTIR ANALYZER SYSTEM
S A W F Reaction/Comment
x x
Notify service.
x x
Notify service.
x x
Check plug connection and wiring.
x x
Check plug connection and wiring.
x x
Check configuration with TCT.
x x
Check plug connection and wiring.
x x
Notify service.
x x
Notify service.
x x
Notify service.
x x
Check measuring gas concentration. Notify service.
x
Check analyzer module and sample preparation. Permis-
sible range: 150% of the smallest measuring range in-
stalled. As soon as the drift exceeds these values, notify
x x
service.
x
Adjust detector displayed manually at the zero point and
end point. Check analyzer module and sample prepara-
tion. Permissible range: 50% of the sensitivity of the
detector. As soon as the drift exceeds this value, notify
x x
service.
x
These messages are created by the automatic adjust-
ment. Check plausibility of the adjustment. Rectify pos-
sible cause of an implausibility. Adjust detector dis-
played manually at the zero point (no. 306) or at the end
x
point (no. 307). Permissible range: 15% of the smallest
measuring range installed; 6% of the smallest measuring
range for measurements on plants and systems requir-
ing approval of 27 and 30 BImSchV
x x
Notify service.
x
see status message of the relevant temperature
x
see status message of the relevant temperature
|
OI/ACF5000-EN REV. A
181