ABB ACF5000 Manual page 73

Multi-component ftir analyzer system
Table of Contents

Advertisement

Overview
Qty.
Adjustment
MC MR method
1
1
Test gas/Single
≥ 1
> 1
Test gas/Common the measuring ranges
> 1
≥ 1
Substitute gas
MC = measuring and substitute gas components
MR = measuring ranges per component
Setting the adjustment method
ACF5000 FTIR ANALYZER SYSTEM
The following table presents the adjustment methods at a glance.
To be configured ...
To be adjusted ...
the zero point and the
end point in every meas-
uring range individually
for every measured
component
the zero point in a meas-
for zero point and
uring range and the end
end point adjustment
point in another measur-
ing range for every
measured component
the components and
the zero point in a meas-
measuring ranges for
uring range of a compo-
zero point and end
nent and the end point in
point adjustment
a measuring range of
another component every
detector
The adjustment method can be set separately for every one of the three
types of controlling the adjustment (manual, automatic and externally con-
trolled).
The measuring ranges for the zero point and end point adjustment for
common and substitute gas adjustment are set jointly for all three types of
controls.
For the substitute gas adjustment, the components need to be additionally
set for the zero point and end point adjustment.
|
OI/ACF5000-EN REV. A
The adjustment affects ...
only on the relevant
measuring range
on all measuring ranges of
the relevant measured
component
on all components and
measuring ranges of the
relevant detector
73