6.2.18 EXECUTE TEST PROGRAM

A test program can contain operator prompts to enter numeric values from instruments; test
results can be stored in the ADTS 405 and later printed, by a suitable printer, using the serial output
port or ethernet option.
6.2.18.1 Executing Test Programs (without auto run)
a.
Press the EXECUTE TEST PROGRAM key. This displays the first of the available test
programs. The top line shows the file name and the second and third lines show the TITLE.
b.
Use [NEXT] or [PREV] to step through the available programs until the required program is
displayed. Press [RUN] to select the program.
The display shows a menu allowing the following selections:
[F1] Run all tests: This starts the program from the beginning.
[F2] Run specific tests: This allows execution to start with any one of the individual TESTS
contained within the test program. Use [NEXT], [PREV] and [SEL] as above.
c.
Select [Run all tests]: The display returns to the normal user display. "TP", in the bottom
right hand corner of the display, indicates execution of a test program. Various commands
prompts the operator by displaying messages on screen such as: "Press the ETP key". The
"ETP" key is the general key for moving onto the next step.
Note: The CLEAR/QUIT key should only be used to terminate test program
execution.
d.
At the end of each test, the operator has a choice of continuing, exiting test program
execution or going to manual mode. In manual mode, any of the normal key-pad functions
may be used.
e.
When the manual testing is complete, press EXECUTE TEST PROGRAM and the test
program continues from where it left off. This feature may be disabled by including the
KEYLOCK command in the test program. While in manual mode the display prompts
"TPM".
f.
When a "go" command is executed, "GO" flashes in the lower right hand corner of the
screen. At this point the user should record any instrument readings or displayed leak
rates, as required, before continuing by pressing EXECUTE TEST PROGRAM. F4 may be used
to continue at a "go" command as an alternative to pressing EXECUTE TEST PROGRAM.
6.2.18.2 Loading Program Files
New program files may be prepared on a PC and transferred to the ADTS using the Ethernet port
access procedure described in section 3.7.2.
6.2.18.3 Deleting program files
If a test program within an ADTS 405 is no-longer required, then it may be deleted. To delete a test
program, proceed as follows:
a.
Enter configuration (CONFIG) by pressing and holding F1 and pressing SETUP.
b.
At the first menu level of configuration, press the EXECUTE TEST PROGRAM key, then F2.
The display then shows the file name and title of the first file in the list. Use [NEXT] and
[PREV] to step through the files until the display shows the file to be deleted.
c.
Select [DEL] to delete the file, then press CLEAR/QUIT to quit from configuration.
d.
Select [YES] to confirm quit.
Note: If the deleted program is marked as "AUTO RUN" then an additional
screen asks to confirm if the auto run status should be removed before deletion.
78 | ADTS 405 Mk2 Instruction Manual-English
© 2015 General Electric Company – All rights reserved.