Communications Panel Options; Ethernet System - GE Druck ADTS 405 R/F Mk2 Instruction Manual

Air data test set
Table of Contents

Advertisement

3.7 Communications panel options

3.7.1 Ethernet System

The ADTS405 is equipped with a TCP/IP Ethernet interface, and appears on the network as up to
two separate IP addresses. The first IP address relates to the ADTS instrument supervisor and
contains access to the functions described above. The second address relates to the
communications adapter (if fitted). There are no user related functions accessible within the
communications adapter IP port at this time, however the functionality of the port may be assessed
and tested if required.
The system is supplied with factory default IP addresses as follows:
192.168.2.2
192.168.2.3
The ports are most simply tested using a network 'ping' to these addresses.
Connect the ADTS to a computer equipped with a network interface, using a UTP network cable.
The network interface should be configured with an IP address on the same subnet.
Example:
If the ADTS has the default IP addresses then the host network address should be configured as
192.168.2.1/255.255.255. The method for doing this may vary but the following example screens
show the method under Windows 7. It is suggested to take careful note of the original computer
settings so they may be restored after the check.
Figure 3-9: Configuring Windows For ADTS Subnet IP Address
A ping to either of the above ADTS IP addresses should give the following typical response seen in
figure 3-10 Ping Test from Windows Command Prompt below:
© 2015 General Electric Company – All rights reserved.
Communications panel options
Supervisor
Communications Adapter
ADTS 405 Mk2 Instruction Manual-English | 45