Config, [Units]; Config, [Limits], [Edit Limits], [Edit Existing] - GE Druck ADTS 405 R/F Mk2 Instruction Manual

Air data test set
Table of Contents

Advertisement

6.2.57.1 Procedure
a.
Holding down F1 then pressing SETUP causes the system to enter CONFIGuration mode.
CONFIGuration may be password protected using a four digit PIN. If PIN protected, the
display prompts for entry of the PIN. Press each digit of the PIN in turn (ENTER is not
required) as each digit is entered, an asterisk appears. If an incorrect PIN is entered, the
complete PIN is requested again.
b.
When the correct PIN is entered, the CONFIG menu appears.
c.
The PIN may be changed using CONFIG,-000.
d.
To use the configuration menu, press the appropriate function key to select the required
menu item. To configure a parameter key, enter configuration menu then press the
required parameter key. Each parameter key configuration provides changes to the normal
user mode function and to the settings available in the set-up menu.
e.
Once a parameter has been changed in configuration, use CLEAR/QUIT to return to the
previous menu. After all required key parameters have been changed, repeatedly press
CLEAR/QUIT until the display shows the normal operating mode.
6.2.57.2 Functions
In the following section, each description of a CONFIGuration function is headed by the menu/key
sequence required to select it. Intermediate menus are not described. Many of the functions are
equivalent to the set-up functions; refer to SETUP for the details of the function.

6.2.58 CONFIG, [UNITS]

This function allows the default units to be configured.
1.
The default [AERONAUTICAL] units are used at power-up for the altitude (and airspeed)
display. The default units are also used for the limits entry in CONFIG and entry of any
altitude or airspeed related functions in SETUP or CONFIG e.g., airfield altitude or altitude
correction.
2.
The default [PRESSURE] units are used for limits entry in CONFIG and entry of any pressure
values in SETUP or CONFIG.
3.
The default [TEMPERATURE] units are used for entry of temperature values in SETUP or
CONFIG.
4.
The default [AERONAUTICAL] and [PRESSURE] are set in MINIMUM SETUP, [UNITS].

6.2.59 CONFIG, [LIMITS], [EDIT LIMITS], [EDIT EXISTING]

This function allows existing sets of limits to be configured.
a.
Use [NEXT] or [PREV] to select the set of limits by aircraft name or component/system
name/number when the display shows the required aircraft limits, press [SEL] to edit the
limit values.
b.
Use [NEXT] or [PREV] to step through each limit or associated parameter in turn. Enter a
new numeric value to change a limit.
c.
Press CLEAR/QUIT when all required changes have been made and select [YES] to save
changes.
© 2015 General Electric Company – All rights reserved.
Introduction
ADTS 405 Mk2 Instruction Manual-English | 91

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents