Set-Up, Rate Timer; Set-Up, Rate; Setup, Leak Measure Control, [Auto Leak]; Setup, Leak Measure Control, [Auto Limit] - GE Druck ADTS 405 R/F Mk2 Instruction Manual

Air data test set
Table of Contents

Advertisement

Chapter 6. REFERENCE AND SPECIFICATION

6.2.42 SET-UP, RATE TIMER

This function allows the Wait/Time values to be set for F1, F2, and F3 of the RATE TIMER function. To
change the times:
a.
Select the function to be changed: F1, F2 or F3.
b.
Enter the [WAIT] value in minutes and seconds.
c.
Enter 00:00 if a wait is not required.
d.
Select [TIME].
e.
Enter the [TIME] value in minutes and seconds.
f.
Press CLEAR/QUIT.
g.
Select another WAIT/TIME or press CLEAR/QUIT.

6.2.43 SET-UP, RATE

This function selects automatic airspeed rate control ON or OFF. If the automatic airspeed rate is
selected (using SETUP, RATE) the airspeed rate is automatically adjusted to simultaneously
achieve both altitude and airspeed aim. This adjustment normally requires a reduction in the rate
from the aim value; the rate does not exceed the aim value. When airspeed is automatically
reduced, the display shows RATE CAS aim as (A).
This function only applies when using Aeronautical units (ft and kts, ft and mph, m and km/h (m/
min), m and km/h (m/s), m and km/h (km/min)).

6.2.44 SETUP, LEAK MEASURE CONTROL, [AUTO LEAK]

This function enables or disables the auto leak recovery facility. See LEAK MEASURE/CONTROL.

6.2.45 SETUP, LEAK MEASURE CONTROL, [AUTO LIMIT]

This function enables or disables the auto limit recovery facility. See LEAK MEASURE/CONTROL.

6.2.46 SETUP, GROUND

This function enables the station (airfield) altitude to be entered and is used when displaying QNH.
See GROUND for further details.
a.
Enter the station (airfield) altitude in the units shown on the display. See CONFIG, [UNITS],
[AERONAUTICAL] for default units.

6.2.47 SETUP, [PORT]

This function sets up the port option. If an option, that is not fitted, is selected the display shows the
error message "Option Hardware not fitted". Refer to the Line Switching Unit, User Manual.

6.2.48 SETUP, PRINT, [DATE/TIME]

This function allows the system time and date to be set. The time and date is maintained by a
battery-backed clock when the ADTS 405 is switched off.
a.
Use [TIME] or [DATE] to select between time or date.
b.
To change the time or date, enter all six digits and press ENTER.
c.
The date format can be either US (mm/dd/yy) or UK (dd/mm/yy) using CONFIG,[MORE],
[DATE FORMAT]. The time format is hh:mm:ss.
88 | ADTS 405 Mk2 Instruction Manual-English
© 2015 General Electric Company – All rights reserved.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents