Fluke 4322 Operator's Manual page 90

Automated pressure calibrator
Table of Contents

Advertisement

4322
Operators Manual
10. Run the test as follows:
11. Fill in the UUT information on the UUT information menu then select  to
12. Select  to close and save the test results or  to print the test
3-36
a. On the run test menu, select  to run the test when ready. The
Controller controls pressure to the first target pressure test point and controls the
pressure with the set control mode set in the test settings.
b. Push
to make a measurement when the measurement indicator shows
Ready .
c. Select . The Controller controls pressure to the next target pressure
test point.
Use  to redo a test point.
d. Push
to make a measurement when the measurement indicator shows
Ready .
e. Repeat steps c. and d. again for the remainder test points.
f.
After all test points are complete, select  to open the test results menu.
open up the run test menu.
The information on this menu is used to populate the calibration report.
results.
To print the test results, connect and turn on the printer. If a printer is
not available at the conclusion of the test, the test results can be viewed
at a later time and printed from the Run Test menu. To see previous test
results, select the test from the Test drop-down menu and select "Past
Results".
Note
Note
Note