To prevent damage to the UUT, make sure to select the correct
size adapter fitting with the correct thread type. Ensure that all
hardware used is rated to adequate working pressure, and that
all equipment is in proper working order (for example, no
cracks or stripped threads).
To prevent injury, only trained personnel should make
connections. Always use standard and/or manufacturer
recommended procedures to connect each device.
4. Wrap the threads of the QC fitting with Teflon tape (if required) then put the fitting
Failure to properly tighten the adapter fitting can result in a leak.
5. Connect the female end of the test hose to the male QC fitting on the UUT.
6. Connect the male end of the test hose to the TEST port on the Controller.
The test hose is not required for every test. In some cases, the adaptor can
be connected to the UUT can be mounted directly into the TEST port.
7. Turn on the Controller.
8. Purge the hose to ensure cleanliness and to prevent contamination. Use the
9. Perform a leak test. Use the instructions in "Leak Test" in Chapter 3. If a leak is
2-12
on the UUT (see Figure 2-9).
Adapter
Figure 2-9. UUT Connection Example
instructions in "Hose Purge" in Chapter 3.
found, correct the leak and attempt the leak test again.
Caution
Warning
Note
UUT
Note
gxj044.eps