Fluke 4322 Operator's Manual page 24

Automated pressure calibrator
Table of Contents

Advertisement

4322
Operators Manual
Item
Spring-Loaded Case
Latches
Intensifier Pressure
Selection Keys
(PRESSURE SET PSI)
SUPPLY Port
Input Power Cord
Connector
Fuse
Power Switch
Vent Valve
(VENT)
OUTLET Port
Intensifier Pressure
Gauge (INTENSIFIER
PRESSURE)
Isolation Valve
(ISOLATION)
1-14
Table 1-6. Intensifier Controls and Indicators (cont.)
Name
Spring-loaded latches used to keep the lid closed and sealed.
Keypad used to select the Intensifier pressure. Keypad selections
are in 1,000 psig increments that range from 1,000 psig to 10,000
psig. Each key is has a status LED indicator that visually shows the
selected boost pressure and system status (see ).
Female quick connect (QC) supply port that connects to the
accessory Nitrogen Cylinder or external nitrogen pressure supply.
Supply must be between (500 to 3,000) psig.
To prevent injury, the supply must be shut off
and depressurized to remove the male supply
line connector from the supply port.
Input connector for the supplied power cord. The power entry module
and power cord are a standard appliance inlet (IEC/EN 60320-1/C14)
with a vertical three-prong arrangement.
Electrical fuse that protects the system from overcurrent. For fuse
replacement instructions, refer to "Fuse Replacement" in Chapter 6.
Electrical power switch to turn on and turn off the unit.
Needle valve to manually discharge system pressure to
atmosphere.
Source pressure can be depleted very quickly if it is not turned
off before venting the system.
Female quick connect (QC) outlet port that connects the Controller.
To prevent injury, the Intensifier must be shut
off and depressurized to remove the male
supply line connector from the outlet port. For
quick disconnects with the system on, isolate
the pressure with the isolation valve and open
the vent valve to remove pressure on the
connector.
Visual Indication of the Intensifier pressure in psig.
Needle valve that shuts-off the pressure supply to the OUTLET
Port. Pressure isolation is used for pressure accumulation
troubleshooting procedures. It is also used to relieve pressure on
the OUTLET port to disconnect the hose.
Function
Caution
Note
Caution