Fluke 4322 Operator's Manual page 111

Automated pressure calibrator
Table of Contents

Advertisement

SENS:PRES:RES[?] [res] or SENSE:PRESSURE:RESOLUTION[?] [res]
Read or set the pressure resolution with res in % of reading as indicated on the front panel.
Range = {1.0000 - 0.0001 %}
Read Example:
SENS:PRES:RES?
Response:
0.0100
Set Example:
SENS:PRES:RES 0.001
SENS:PRES:MODE[?] [mode] or SENSE:PRESSURE:MODE[?] [mode]
Read or set the pressure measurement mode where 'mode' can be set to ABSOLUTE, GAUGE, or
VACUUM.
Read Example:
SENS:PRES:MODE?
Response:
ABSOLUTE
Set Example:
SENS:PRES:RES GAUGE
SENS:PRES:RANG? [range] or SENS:PRESSURE:RANGE? [range]
Read the range of the currently active sensor or select which sensor is currently active in the
current units and measurement mode. The active sensor is the only sensor that can be selected
only when it is idle (not generating pressure) and the pressure is within the specified range of the
sensor. The Controller will automatically override a manually specified sensor selection when a
new target is set or if pressure increased above the specified range of the sensor.
Read Example:
SENS:PRES:RANGE?
Response:
1500.0
Set Example:
SENS:PRES:RANGE 50
100.0 (range of most suitable sensor for range request is replied)
Response:
SENS:PRES:REF:HEIG[?] [head] or
SENSE:PRESURE:REFERENCE:HEIGHT[?] [head]
Read or set the height difference between the Controller and the UUT used for the head-height
correction. head is in the current length units and is relative to the Controller.
Read Example:
SENS:PRES:REF:HEIG
Response:
10.0
Set Example:
SENS:PRES:REF:HEIG 3
SENS:PRES:REF:MED[?] [gas] or SENSE:PRESSURE:REFERENCE:MEDIUM[?] [gas ]
Read or set the gas medium used for the head correction between the Controller and the UUT.
'gas' is either N2, HE, or AIR.
Read Example:
SENS:PRES:REF:MED
Response:
N2
Set Example:
SENS:PRES:REF:MED AIR
Remote Operation
Remote Operation Commands
5
5-13