4322
Operators Manual
CALx:PRES:DATE[?] [yyyy,mm,dd] or CALIBRATEx:PRESSURE:DATE[?] [yyyy,mm,dd]
Read or set the user calibration date for the specified sensor. The entered values are all numeric
and "yyyy" is a four digit year (2000-2135). The default value is 2000. "mm" is a two digit month
(1-12), and "dd" is a two digit day (1-31). This date should be set whenever the calibration
coefficients are modified using the
always be used with this command to: (1) ensure that the execution is complete and (2) to make
sure a response has been sent before anther command is sent.
These values are stored within the nonvolatile memory inside the sensor, so this command should
be used cautiously. Suffix 'x' specifies the sensor to access:
Sensor
Read Example:
Response:
Set Example:
CALx:PRES:SER? or CALIBRATEx:PRESSURE:SERIAL?
Read the serial number of the sensor. Suffix 'x' specifies the sensor to access:
Sensor
Read Example:
Response:
CALC:LIM:LOW[?] [limit] or CALCULATE:LIMIT:LOWER[?] [limit]
Read or set the lower limit for requesting set-points in the current displayed pressure units.
Pressure generation will abort and an alarm will sound if the pressure is below this limit.
Read Example
Response:
5.00
Set Example:
5-10
x:
1
10 kpsig
CAL2:PRES:DATE?
2010,05,24
CAL2:PRES:DATE 2011,12, 30
x:
1
10 kpsig
CAL3:PRES:SER?
01987041
: CALC1:LIM:LOW?
CALC:LIM:LOW 110
command. The query operator should
CALx:PRES:DATA
2
3
1500 psig
100 psig
2
3
1500 psig
100 psig
4
5
21.5 psia
Barometer
4
5
21.5 psia
Barometer