SYST:KLOC[?] [0|1] or SYST:KLOCK[?] [0|1]
Locks or unlocks the system keypad (except for the ABORT key). The keypad is always
unlocked when the Controller is powered up. 'n' = 0/1 for unlock/lock.
Read Example:
Response:
Set Example:
SYST:TIME[?] [hh,mm,ss] or SYSTEM:TIME[?] [hh,mm,ss]
Read or set the System Time (24 hour time format only).
Range: hh = {0 to 23}, mm = {0 to 59}, ss = {0 to 59}
Response:
23,51,05
Set Example:
SYST:TIM:COUN? or SYSTEM:TIMER:COUNT?
Gets the elapsed seconds since the unit has been turned on.
Read Example:
Response:
SYST[x]:VER:THM? or SYSTEM[x]:VERSION:THM?
Read the firmware version installed in the sensor.
Specifies the sensor to access: x: 1 2 3 4
Sensor 10 kpsig 1500 psig 100 psig 21.5 psia
Example:
SYST4:VER:THM?
Response:
UNIT:PRES[?] [unit] or UNIT:PRESSURE[?] [unit]
Read or set the pressure units displayed on the UI and one the remote interface. Please see the
user interface section of the Operators Manual for the supported units.
Read Example:
Response:
psi
Set Example:
UNIT:LENG[?] [unit] or UNIT:LENGth[?] [unit]
Read or set the head height units. Possible arguments are "cm" or "in".
Read Example:
Response:
in
Set Example:
UNIT:TEMP[?] [n] or UNIT:TEMPERATURE[?] [n]
Read or set the display temperature units, where "n" is "C" or "F".
Default: C.
Read Example:
Response:
C
Set Example:
5-16
SYST:KLOCK?
1
SYST:KLOC 1
SYST:TIME 14,15,05
SYST:TIM:COUN?
2398
0.19
UNIT:PRES?
UNIT:PRES kPa
UNIT:LENG?
UNIT:LENG cm
UNIT:TEMP?
UNIT:TEMP F