Fluke 4322 Operator's Manual page 86

Automated pressure calibrator
Table of Contents

Advertisement

4322
Operators Manual
Item
Name/Button
Sequence
Window
Max
Min
Step
Direction
6. Set the control mode to use for the test (see "Pressure Control Modes" on page 3-8).
Each test has a control mode setting. When the test is run, the Controller
uses the Pressure Control mode in the test and overrides the Pressure
Control mode set in the general Controller settings. After the test is
complete, the Controller changes back to the Control Mode set in the
general Controller settings.
3-32
Table 3-11. Edit Test Menu (cont.)
Shows the number and target value of each test point.
Add a test point to the test.
Add a vent to atmosphere test point. An ATM test point vents all pressure to atmosphere to
allow the user to make a vented atmosphere measurement.
ATM is used only in the Absolute Measurement mode.
Delete a selected test point from the test.
Delete all test points from the test.
Maximum target pressure of the test. Value is limited to the tolerance of the UUT set in the
new test menu.
Minimum target pressure of the test. Value is limited to the tolerance of the UUT set in the
new test menu.
Divisor percentage that sets the number of test points in a test. The Controller divides the
maximum target value by the span percentage to give the number of test points and target
value for each test point. For example, if the maximum target value is set to 150 psi and the
step is set to 20 %, then the Controller makes six test points in 30 psi increments.
Direction of the test. The four choices are:
Up – Make a set of test points with the first test point at the minimum target value and the
last lest point at the maximum target value.
Down – Make a set of test points with the first test point at the maximum target value and
the last lest point at the minimum target value.
Up/Down – Make a set of test points with the first test point at the minimum target value,
the test steps up to the maximum target value then back down to the minimum target value.
Down/Up – Make a set of test points with the first test point at the maximum target value,
the test steps down to the minimum target value then back up to the maximum target value.
c. Push  to go to the next page of the edit test menu.
Function
Note
Note