Fluke 4322 Operator's Manual page 110

Automated pressure calibrator
Table of Contents

Advertisement

4322
Operators Manual
MEAS:PRES:SLEW? or MEASURE:PRESsure:SLEW?
Reads the slew rate for the active sensor. The response is sent in the current pressure
measurement unit per second when the next measurement is available.
Example:
MEAS:PRES:SLEW?
Response:
MEAS[x]:ZERO:VAL[?] [P
Read or set the zero values for the three measurement elements (A, B, and C) within the specified
sensor. These values are in the selected units in full resolution and are determined and set by the
autozero routine when autozero is enabled and the Controller vented and stable. The suffix 'x'
specifies the sensor to access:
Sensor
Read Example:
Response:
Set Example:
MEAS:ZERO:STAT[?] [0|1] or MEASURE:ZERO:STATE[?] [0|1]
Read or set the autozero state. When set to '1', autozero will be enabled and will zero the sensors
when the Controller is vented and stable. When set to '0', autozero will be disabled and the zero
values for the three measurement elements (A, B, and C) will not be updated. Autozero is always
enabled on power up regardless of its prior state.
Read Example:
Response:
Set Example:
OUTP:STAT[?] [0|1] or OUTPUT:STATE[?] [0|1]
Read or set the current operation state of the Controller. A "0" idles the Controller for
measurement use only. '1' resumes control at the last specified set-point. The response to the
query form indicates if the controller is active ('1') or idle ('0').
Read Example:
Response:
0
Set Example:
OUTP:MODE[?] [mode] or OUTPUT:MODE[?] [mode]
Read or set the current controller operating mode. 'mode' is the selected mode:
Mode
MEASURE
CONTROL
VENT
Read Example:
Response:
VENT
Set Example:
5-12
0.28
,P
,P
Z,A
Z,B
x:
1
10 kpsig
MEAS:ZERO:VALUE?
-16105.5,-16580.6,-21517.1
MEAS:ZERO:VALUE 0,0,0
MEAS:ZERO:STAT?
1
MEAS:ZERO:STAT 0
OUTP:STAT?
OUTP:STAT 1
Idles Controller until a new set-point is specified or control is restarted.
OUTP:MODE?
OUTP:MODE MEAS
] or MEASURE[x]:ZERO:VALUE[?] [P
Z,C
2
3
1500 psig
100 psig
Description
Restart control at the last specified set-point.
Vents the Controller.
,P
,P
]
Z,A
Z,B
Z,C
4
21.5 psia