3. Use the keys to enter the timing data in the input fields.
4. Set the "Mode" to "Active".
5. Apply the entered data by selecting "Execute function..."
Note
Always save the data with "Execute function...".
The timing data are only entered into memory with the "Execute function..." command, and
the image is then displayed with the new parameters.
6.6.4 Determining the timing data
If you know none or only part of the timing data, you can determine the data incrementally in
the "Force Mode" menu.
Note
All settings must be implemented with the zoom factor "1 to 1" ("Geometry" menu).
Sequence for determining a timing
1. Determine the start values using the auto function.
2. Enter the known data.
3. Optimize the scanning frequency.
4. Optimize the geometry and resolution.
Step-by-step instructions for setting a timing can be found in the next sections.
CuratOR LX491W
Instructions for Use, 02/2018
6.6 Adapting the monitor using Force Mode
Commissioning
29