Instructions for Use 4MP 30" LCD Monitor Important Please read the safety information and all information delivered with the product carefully to familiarize yourself with safe and effective usage.
Use of EIZO products WARNING EIZO products may only be used for the applications described in the catalog and in the relevant tech- nical documentation. If products and components from other manufacturers are used, these must be recommended or approved by EIZO. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
Description of OSD menu .................. 27 7.3.1 Display menu .................... 28 7.3.2 Picture Layout menu ................. 30 7.3.3 Power Manager menu................ 32 7.3.4 Other Options menu.................. 33 7.3.5 Information menu .................. 34 7.3.6 Signal menu .................... 34 RadiForce LX300W Instructions for Use, 07/2019...
Page 4
13.2 Markings and Symbols ................... 51 13.3 Environmental protection .................. 52 13.4 China RoHS (Restriction of Hazardous Substances) .......... 52 13.5 FCC Declaration of Conformity................ 54 13.6 Additional devices.................... 54 13.7 Repairs ........................ 54 13.8 Contact ........................ 55 13.9 Trademarks...................... 55 Index .......................... 56 RadiForce LX300W Instructions for Use, 07/2019...
1 Introduction 1.1 Contents of this document This document explains the functionality and the approved use of the RadiForce LX300W. To ensure clarity, it does not contain all detailed information on this product. The contents of this document are neither part of a previous or existing agreement, commit- ment or legal relationship, nor does it modify such.
2 Safety information 2.1 General safety instructions Correct and safe operation of EIZO devices assume professional transport, storage, installa- tion, and connection, as well as careful operation and service. The devices may only be used for applications for which they are intended.
Page 7
Only use power cables approved for China. These power cables are identified by the labels "CCC" or "CQC". Observe the country-specific regulations Observe all regulations of the country in which the device is used. RadiForce LX300W Instructions for Use, 07/2019...
Page 8
If the device is brought into a room with a higher or rising temperature, condensed water will form in and on the device. Do not switch on the device until the condensed water has evaporated. Other- wise, the device could be damaged. RadiForce LX300W Instructions for Use, 07/2019...
Do not connect devices which are not part of the medical system. NOTICE Opening the device The device must only be opened by service personnel. • Disconnect the power supply plugs before opening the device RadiForce LX300W Instructions for Use, 07/2019...
Page 10
Note Image quality To maintain constant image quality, EIZO recommends cleaning the monitor on a regular basis and checking image properties in accordance with all applicable local regulations. RadiForce LX300W Instructions for Use, 07/2019...
Keep the packaging material for subsequent transport of the device. Device The RadiForce LX300W is a 4MP 29.8" LCD Monitor for installation in a ceiling or wall mount, or for installation in a stand. The RadiForce LX300W-P version has protective glass and the RadiForce LX300W-S version a stand.
Page 12
The LX300W is precalibrated at the factory. A total of five practice oriented Look Up Tables (LUTs) have been preset. This calibration data makes installation and maintenance easier. As such, the monitor can be easily adapted to the respective application and local lighting conditions. RadiForce LX300W Instructions for Use, 07/2019...
The following presents the screen layouts that can be selected in the OSD menu, with spec- ification of the maximum resolution and possible connections. Screen layout for one video source Screen layout for two video sources (PaP) RadiForce LX300W Instructions for Use, 07/2019...
Page 14
Description 3.3 Monitor operating modes Screen layout for two video sources (PiP) Screen layout for three video sources Screen layout with fail-safe function RadiForce LX300W Instructions for Use, 07/2019...
In the worst case, deposits are possible which become evident as dark spots in a white picture and result in deterioration of the luminance. • Protect the monitor from dust, for example through appropriate construction measures at the installation site. RadiForce LX300W Instructions for Use, 07/2019...
Holders must be tested and approved by the manufacturer for the weight to be sup- ported. • An installed stand must be sufficiently stable such that tilting up to 10° does not result in the monitor toppling. RadiForce LX300W Instructions for Use, 07/2019...
Device settings may only be adjusted by service personnel. NOTICE Disconnecting from line power Always set the power switch to "Off" before disconnecting the device from power. Other- wise the device could be destroyed. RadiForce LX300W Instructions for Use, 07/2019...
DVI - single link connection (DVI 2) ⑪ Power connection ⑥ Display port connection ⑫ Power switch 5 V connections The monitor has two 5 V outputs which can be used to supply external devices. RadiForce LX300W Instructions for Use, 07/2019...
Page 19
• Only photometers tested for calibrating the monitor may be connected to the serial inter- face. • A photometer must not be connected in the presence of the patient. Service connection The service connection is used by Service for software updates. RadiForce LX300W Instructions for Use, 07/2019...
Do not use any other fuse. CAUTION Connection in the USA and Canada Molded power plugs must comply with the requirements for "hospital grade attachments" CSA Std. C22.2 No. 21 and UL 498. RadiForce LX300W Instructions for Use, 07/2019...
• DVI 2 can only process single link signals. • Only use the dual link DVI cables specified by EIZO or the transmission links supplied by EIZO. The picture quality, interference resistance, and electromagnetic interference of the en- tire system depend on the length and quality of the cables.
– First, insert the latch on the top of the cover on the monitor. – The press the latches to the side again. 7. If necessary, secure the device with a Kensington Lock at the Kensington lock opening. RadiForce LX300W Instructions for Use, 07/2019...
Use a screen saver with regularly changing images • Switch off the monitor when it is no longer needed. • The monitor has an energy saving mode: If the application in use supports the energy saving mode, activate it. RadiForce LX300W Instructions for Use, 07/2019...
2. Once the driver has been installed successfully, the resolution can be configured using the driver- or operating system settings. Note Installation and parameterization of the graphics card Please refer to the graphics controller manufacturer’s manual for detailed information about installation and configuration of the graphics controller. RadiForce LX300W Instructions for Use, 07/2019...
④ Up (up arrow) ⑤ Enter ⑥ Power ⑦ Operation LED Control panel The LX300W has a touch-sensitive control panel. The keys are located at the lower center of the monitor on the front. RadiForce LX300W Instructions for Use, 07/2019...
2. Then press the "Down" key ③ three times. The LED display brightness changes in response when the keys are pressed. The OSD menu is now locked or unlocked, depending on its initial state. RadiForce LX300W Instructions for Use, 07/2019...
The OSD menu is used to make settings for operation of the monitor with up to three video sources. The layout of the OSD menu is shown in the following figure. Fig.: OSD menu layout RadiForce LX300W Instructions for Use, 07/2019...
Change the brightness of the backlight Default: 800 You change the backlight brightness here. If you adjust brightness, the gay scale values no longer correspond to the set DICOM curve. The calibrated values therefore cannot be guaranteed. RadiForce LX300W Instructions for Use, 07/2019...
Page 29
DVI 3 input is displayed in the upper left corner. Note: Fail Safe can be switched on only when Single Source or Dual Source (PaP) is set in Picture Layout. RadiForce LX300W Instructions for Use, 07/2019...
Note: If Fail Safe is active, the Dual Source (PiP) or Triple Source function cannot be selected. De- activate Fail Safe in the "Display" OSD menu. RadiForce LX300W Instructions for Use, 07/2019...
Page 31
Set To Aspect: The picture is zoomed to Default: 1:1 the maximum window area with retention of the aspect ratio. • Fill All: The image fills the entire area of the PiP window. The aspect ratio is changed. RadiForce LX300W Instructions for Use, 07/2019...
"Backlight Dimmed". Prior to diagnosing radiographic images however, the lamp requires a warm-up time after reactivation. • Disabled: DMPM signals are ignored. The monitor does not change to energy saving mode. RadiForce LX300W Instructions for Use, 07/2019...
If the values do not differ from the defaults, "Nor- mal" status is displayed. Reset To Factory Defaults Selecting this function opens a dialog box where you can reset the device to the factory settings. RadiForce LX300W Instructions for Use, 07/2019...
7.3.6 Signal menu Indicator Description Input Selecting this function displays information regarding the video signals at the inputs. Width (Pixels) #### The inputs displayed by the monitor are highlighted in gray. Height (Lines) #### RadiForce LX300W Instructions for Use, 07/2019...
Fairy Ultra, Pril, Palmolive liquid Pyridine derivatives Activ spray, undiluted Water Tap water Distilled water Note Information on cleaning or disinfection of other system components can be obtained from the respective instructions for use. RadiForce LX300W Instructions for Use, 07/2019...
The picture quality of the monitor changes due to aging of the LCD unit and the backlight. • Check the monitor settings at regular intervals in accordance with the local guidelines. • Correct the settings if necessary. RadiForce LX300W Instructions for Use, 07/2019...
Device is off • Switch on power switch Power cable is not inserted or incor- • Check the power cable rectly inserted. Power cable is defective • Replace power cable Blown fuse • Inform service department RadiForce LX300W Instructions for Use, 07/2019...
Locking or unlocking the OSD menu [} 26] OSD Unlocked Unlocking the OSD menu was suc- cessful Sensor Status Normal Blacklight sensor works problem-free Wrong OSD version The wrong OSD version is installed in Contact service. the monitor. RadiForce LX300W Instructions for Use, 07/2019...
AC 100 ... 240 V (± 10 %) Line frequency 50 ... 60 Hz (± 5 %) Current consumption max. 1.5 A ... max. 0.7 A Maximum current consumption 150 W Energy saving mode < 20 W Standby mode Typically 1 W RadiForce LX300W Instructions for Use, 07/2019...
IP20 according to EN 60529 Connector panel On rear panel, under cover Weight (without stand) RadiForce LX300W: 15.5 kg +/- 1 kg RadiForce LX300W-P: 18.5 kg +/- 1 kg RadiForce LX300W-S: 19.5 kg +/- 1 kg Dimensions (W x H x D) in mm 701.5 x 472.6 x 107.4...
CAN/CSA - C 22.2 No. 601.1-M90 • GB4943.1 (non-tropical, altitude < 2000 m), safety specification T8AL, 250 V Protection class Protection class I Degree of protection IP20 Medical device classification Class I according to 93/42/EEC Appendix IX RadiForce LX300W Instructions for Use, 07/2019...
Dimension drawings 11.1 View from the front and above 11 Dimension drawings All dimensions in mm 11.1 View from the front and above RadiForce LX300W Instructions for Use, 07/2019...
Transmission over Ethernet cables permits robust handling and easy installation. The cable can also be easily routed through small openings. Order no.: 6GF6010‑2DB36 DVI-D Dual Link signal cable EIZO recommended DVI-D Dual Link signal cable, 2 m in length. Order no.: 6GF6980-1TA10 RadiForce LX300W Instructions for Use, 07/2019...
IEC/EN 60601-1-2. Electromagnetic radiation The RadiForce LX300W is intended for use in the electromagnetic environments noted below. Customers and users of the RadiForce LX300W have to ensure that the device is used in such an environment. Radiation test Conformity...
Page 47
The RadiForce LX300W was tested with the following compliance levels in accordance with the test requirements for professional healthcare facilities, as established in IEC/EN 60610-1-2. Customers and users of the RadiForce LX300W have to ensure that the monitor is used in such an environment.
Page 48
13.1 Information on electromagnetic compatibility (EMC) Electromagnetic interference immunity The RadiForce LX300W was tested with the following compliance levels in accordance with the test requirements for professional healthcare facilities, as established in IEC/EN 60601-1-2. Customers and users of the monitor have to ensure that the monitor is used in such an environment.
Page 49
Recommended minimum distance between portable or mobile RF communications devices and the RadiForce LX300W The RadiForce LX300W is intended for use in an electromagnetic environment in which interference due to electromagnetic radiation is controlled. For other portable and mobile RF communication de- vices (transmitters), the recommended minimum distance between the portable and mobile RF com- munication devices (transmitters) and the monitor applies as listed below.
Page 50
Recommended minimum distance between portable or mobile RF communications devices and the RadiForce LX300W The RadiForce LX300W is intended for use in an electromagnetic environment in which interference due to electromagnetic radiation is controlled. The customer or user of the monitor can help prevent electromagnetic interference by maintaining the recommended minimum distance between portable and mobile RF communications devices (transmitters) and the monitor.
Symbol for degree of protection according to DIN EN 60529. "On" symbol (voltage). "Off" symbol (voltage). Input for service calls. Symbol for "Observe operating instructions". Symbol for "Authorized service personnel only". Symbol for "Hazardous electrical voltage". RadiForce LX300W Instructions for Use, 07/2019...
EFUP shall be done in accordance with the Product Maintenance Procedures. This product must not be disposed of as unsorted municipal waste, and must be col- lected separately and handled properly after decommissioning. RadiForce LX300W Instructions for Use, 07/2019...
Page 53
Applications of hazardous substances in this medical device are required to achieve its intended clinical uses, and/or to provide better protection to human beings and/or to environment, due to lack of reasonably (economically or technically) available substitutes. 产品中有毒有害物质或元素的名称及含量 Table of hazardous substances’ name and con- centration. RadiForce LX300W Instructions for Use, 07/2019...
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note Use the attached specified cable below or EIZO signal cable with this monitor so as to keep interfer- ence within the limits of a Class B digital device. •...
13.9 Trademarks The EIZO Logo is a registered trademark of EIZO Corporation in Japan and other countries. EIZO is a registered trademark of EIZO Corporation in Japan and other countries. RadiForce is a registered trademark of EIZO Corporation in Japan and other countries.
Service personnel 5 Shielding measures 17 Switch on 23 Image Sticking 23 Installation site 15 Intended use 5 Technical specifications 39 Troubleshooting 37 Landscape format 29 Line power 20 USB connections 20 User 5 RadiForce LX300W Instructions for Use, 07/2019...
Page 57
Index Video source Window arrangement 31 Zoom 31 Window Configuration 31 RadiForce LX300W Instructions for Use, 07/2019...