Ti Interface - Renishaw ATOM Installation Manual

Linear encoder systems
Hide thumbs Also See for ATOM:
Table of Contents

Advertisement

System connection:

Ti interface

Approved ESD precautions must be followed
at all times during readhead and interface
electrical connection. The readhead is
connected to the Ti interface via a small,
rugged interboard connector to allow for
easy feed-through during installation.
Connecting the readhead
1
Remove the cover plate as shown
(2 off M2.5 hex head screws).
2
Taking care not to touch the pins,
plug the connector into the
socket in the interface,
ensuring correct orientation as shown.
3
Refit the cover plate ensuring the cable ferrule is located in the recess on the inside and no wires are trapped under the cover plate.
4
When the system has been connected proceed with 'Readhead mounting and alignment' and 'System calibration' sections.
ATOM
linear installation guide
Disconnecting the readhead
1
Disconnect power.
2
Remove the cover plate on the interface
(2 off M2.5 hex head screws).
3
Gently lever the connector PCB
(on the end of the cable) out of the socket.
Do not pull the cable to remove the connector.
4
Place the connector in an anti-static bag.
5
Refit the cover plate.