Dissolved oxygen
30

5.3.5 Calibration with Comparison meas. (FDO Comp)

For this calibration procedure, the Comparison meas. setting must be set to on
in the Calibration menu.
Before calibrating via a comparison measurement, the sensor
should be calibrated in the air calibration vessel.
Proceed as follows to calibrate the instrument:
1.
Connect the D.O. sensor to the meter.
2.
Immerse the D.O. sensor in the reference solution.
3.
Start the calibration with <CAL>.
O
2
FDO Comp
4.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The multiplier that was set last is displayed.
O
2
FDO Comp
7.92
mg/l
25.1
°C
22.09.2008
08:00
7.92
mg/l
0.973
22.09.2008
08:00
Oxi 3310 IDS (Oxi 3315)
ba75944e02
10/2014