Sensor-Independent Settings; System - wtw Oxi 3310 IDS Operating Manual

Digital meter for digital ids-d.o. sensors
Table of Contents

Advertisement

Settings
Menu item
System / General / Lan-
guage
System / General / Audio
signal
System / General / Illumi-
nation
System / General / Con-
trast
System / General / Shutoff
time
System / General / Tem-
perature unit
34
Menu item
Salinity
Response time t90
Reset
6.2

Sensor-independent settings

6.2.1 System

To open the Storage & config menu, press the <ENTER_> key in the measured
value display. After completing the settings, switch to the measured value dis-
play with <M>.
Default settings are printed in bold.
Possible setting
Deutsch
English
(more)
on
off
Auto
on
off
0 ... 50 ... 100 %
10 min ... 1h ...
24 h
°C
°F
Possible
Explanation
setting
0.0 ... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item if only available if the
manual salinity correction (Sal cor-
rection) isswitched on.
30 ... 300 s
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces
the limits of variation of the mea-
sured value. The signal filter is char-
acterized by the response time t90.
This is the time after which 90 % of a
signal change is displayed.
-
Resets all sensor settings to the
delivery condition (see section 6.3.1
R
ESETTING THE MEASUREMENT SET
TINGS
Explanation
Selects the menu language
Switches on/off the beep on keystroke
Switches the display illumination on/off
Changes the display contrast
Adjusts the switch-off time
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperature values are displayed with the
selected unit.
Oxi 3310 IDS (Oxi 3315)
, page 37).
ba75944e02
-
10/2014