18. FAQ
Where do I find important instru-
ment information?
The instrument shows the error
message „O2 sensor 0 mV"!
The instrument shows the mes-
sage „Check required"!
The instrument shows the error
message „T-Gas" or „T-Air"!
The instrument shows wrong or
inaccurately CO
values!
2
My instrument cannot be
switched on!
My instrument does not print!
Can I change the printout?
pro
ecom-
J2KN
In the menu „Control" all important instrument
information are shown (e.g. battery voltage,
sensor values, unit number, next service date,
operation hours etc.). With the arrow keys
stands you can switch to the second page.
The sensor must be renewed.
This message appears automatically every 12
months. Note: This is a recommendation to let
check the instrument. The instrument is how-
ever still ready for use.
Possible reasons could be:
- Cable is broken (at the plug)
- T-Air sensor is broken
- Thermocouple is broken
- Cable is defective
Note: The error messages can be ignored at
the J2KN
tions that depend on these temperatures are
not implemented.
Possible reasons could be:
- O
- Pump is not working correctly
- Leakage in the gas way
- Condensate trap / gas cooler is clogged
- Please check the mains cable
- Please check the fuse
- Please check mains connection (Plug socket
- Please load the accumulator min. 8 hours
Please check whether the printer paper is
correctly inserted. The thermal printer writes
only on the thermally sensitive side. Please
use always the correct paper for the printer,
you will prevent defects at the printer. Please
make sure that the printer is clean (no chads
in the drive).
You can change the printout (Menu: Adjust-
ments).
TECH
pro
TECH by pressing „OK". Calcula-
is defective (CO
2
from the O
values)
2
switched on?)
(Accumulator could be over-discharged)
values are calculated
2
Page 53