Start Trigger Character (Non-Delimited)

Usage:
Useful in applications where different characters are required to start a
read cycle.
Definition:
A single ASCII host serial trigger character that starts the read cycle and is
not enclosed by delimiters such as
Non-delimited Start characters can be defined and will function according
to the trigger event.
When defining Start trigger characters, the following rules apply:
• In External Edge the MS-1 looks only for the Start trigger character
• In External Level the Start trigger character begins the read cycle and
• In Serial Data and Edge trigger mode, either a Start trigger character
Serial Cmd:
<K229,start trigger
Default:
NUL
Options:
Two hex digits representing any ASCII character except XON and XOFF.
MS-1 Scan Engine User Manual
and ignores any Stop trigger character that may be defined.
the Stop trigger character ends it. Note that even after a symbol has
been decoded and the symbol data transmitted, the MS-1 remains in
External Level trigger read cycle until a Stop character is received.
or a hardware trigger can start an edge trigger read cycle.
character>
(00 in hex) (disabled)
<
and >.
Read Cycle
6-11